What Refractometer is all about?

 

A refractometer is a simple instrument used for measuring concentrations of aqueous solutions. It requires only a few drops of liquid, and is used throughout the foodagriculturalchemical, and manufacturing industries.

 

Modern refractometers

How a Refractometer Works?

When light enters a liquid it changes direction; this is called refraction. Refractometers measure the degree to which the light changes direction, called the angle of refraction. A refractometer takes the refraction angles and correlates them to refractive index (nD) values that have been established. Using these values, you can determine the concentrations of solutions. For example, solutions have different refractive indexes depending on their concentration in water.

           

A Refractive index
Brix Scale and Common Brix %
The prism in the refractometer has a greater refractive index than the solution. Measurements are read at the point where the prism and solution meet. With a low concentration solution, the refractive index of the prism is much greater than that of the sample, creating a large refraction angle and a low reading (Table 1). The reverse would happen with a high concentration solution (Table 2 ).

The Brix scale is calibrated to the number of grams of cane sugar contained in 100 mL of water. Therefore, the Brix % reading equals actual sugar Concentration.

Sample fluid Brix %
Cutting oils
Oranges
Carbonated beverages
Apples
0 to 8
4 to 13
5 to 15
11 to 18
Grapes and wines
Concentrated juices
Condensed milk
Jams and jellies
14 to 19
42 to 68
52 to 68
60 to 70

Table 1 showing the Brix %

Common Refractive Indexes

Sample fluid Temperature Refractive
index
Methanol
Acetone
Ethanol
Acetic acid
Benzene
25°C
25°C
25°C
25°C
25°C
1.326
1.357
1.359
1.370
1.498
Paraffin oil
Palm oil
Olive oil
20°C
20°C
20°C
1.412
1.456
1.471
Methyl salicylate
Methyl iodide
25°C
25°C
1.522
1.740

Table 2 : Refractive index readings are temperature-dependent.

TYPES OF REFRACTOMETERS

There are four main types of refractometers:

There is also the Rayleigh Refractometer used (typically) for measuring the refractive indices of gases.

USES OF A REFRACTOMETER

In veterinary medicine, a refractometer is used to measure the total plasma protein in a blood sample and urine specific gravity in a urine sample.

In drug diagnostics, a refractometer is used to measure the specific gravity of human urine.

In gemology/gemmology, the gemstone refractometer is one the fundamental pieces of equipment used in a gemological laboratory. Gemstones are transparent minerals and can therefore be examined using optical methods. Refractive index is a material constant, dependent on the chemical composition of a substance. The refractometer is used to help identify gem materials by measuring their refractive index, one of the principal properties used in determining the type of a gemstone. Due to the dependence of the refractive index on the wavelength of the light used (i.e. dispersion), the measurement is normally taken at the wavelength of the sodium line D-line (NaD) of ~589 nm. This is either filtered out from daylight or generated with a monochromatic light-emitting diode (LED). Certain stones such as rubies, sapphires, tourmalines and topaz are optically anisotropic. They demonstrate birefringence based on the polarisation plane of the light. The two different refractive indexes are classified using a polarisation filter. Gemstone refractometers are available both as classic optical instruments and as electronic measurement devices with a digital display.

In marine aquarium keeping, a refractometer is used to measure the salinity and specific gravity of the water. In the Automobile Industry, a refractometer is used to measure the Coolant Concentration and the Ph Value of the Coolant Oils for the CNC Machining Process.

In homebrewing, a refractometer is used to measure the specific gravity before fermentation to determine the amount of fermentable sugars which will potentially be converted to alcohol.

In beekeeping, a refractometer is used to measure the amount of water in honey.

AUTOMATIC REFRACTOMETERS

Automatic refractometers automatically measure the refractive index of a sample. The automatic measurement of the refractive index of the sample is based on the determination of the critical angle of total reflection. A light source, usually a long-life LED, is focused onto a prism surface via a lens system. An interference filter guarantees the specified wavelength. Due to focusing light to a spot at the prism surface, a wide range of different angles is covered. As shown in the figure “Schematic setup of an automatic refractometer” the measured sample is in direct contact with the measuring prism. Depending on its refractive index, the incoming light below the critical angle of total reflection is partly transmitted into the sample, whereas for higher angles of incidence the light is totally reflected. This dependence of the reflected light intensity from the incident angle is measured with a high-resolution sensor array. From the video signal taken with the CCD sensor the refractive index of the sample can be calculated. This method of detecting the angle of total reflection is independent on the sample properties. It is even possible to measure the refractive index of optically dense strongly absorbing samples or samples containing air bubbles or solid particles . Furthermore, only a few microliters are required and the sample can be recovered. This determination of the refraction angle is independent of vibrations and other environmental disturbances.

       Schematic setup of an automatic refractometer

Influence of wavelength

The refractive index of a given sample varies with wavelength for all materials. This dispersion relation is nonlinear and is characteristic for every material. In the visible range, a decrease of the refractive index comes with increasing wavelength. In glass prisms very little absorption is observable. In the infrared wavelength range several absorption maxima and fluctuations in the refractive index appear. To guarantee a high quality measurement with an accuracy of up to 0.00002 in the refractive index the wavelength has to be determined correctly. Therefore, in modern refractometers the wavelength is tuned to a bandwidth of +/-0.2 nm to ensure correct results for samples with different dispersions.

A  modern refractometer with bandwidth of +/-0.2 nm

Influence of temperature

Temperature has a very important influence on the refractive index measurement. Therefore, the temperature of the prism and the temperature of the sample have to be controlled with high precision. There are several subtly-different designs for controlling the temperature; but there are some key factors common to all, such as high-precision temperature sensors and Peltier devices to control the temperature of the sample and the prism. The temperature control accuracy of these devices should be designed so that the variation in sample temperature is small enough that it will not cause a detectable refractive-index change.

External water baths were used in the past but are no longer needed.

A refractometer with external water bath

FLOW CELLS

There are different types of sample cells available, ranging from a flow cell for a few microliters to sample cells with a filling funnel for fast sample exchange without cleaning the measuring prism in between. The sample cells can also be used for the measurement of poisonous and toxic samples with minimum exposure to the sample. Micro cells require only a few microliters volume, assure good recovery of expensive samples and prevent evaporation of volatile samples or solvents. They can also be used in automated systems for automatic filling of the sample onto the refractometer prism. For convenient filling of the sample through a funnel, flow cells with a filling funnel are available. These are used for fast sample exchange in quality control applications.

                

                  An automatic refractometer equipped with a flow cell

Once an automatic refractometer is equipped with a flow cell, the sample can either be filled by means of a syringe or by using a peristaltic pump. Modern refractometers have the option of a built-in peristaltic pump. This is controlled via the instrument‘s software menu. A peristaltic pump opens the way to monitor batch processes in the laboratory or perform multiple measurements on one sample without any user interaction. This eliminates human error and assures a high sample throughput.

If an automated measurement of a large number of samples is required, modern automatic refractometers can be combined with an automatic sample changer. The sample changer is controlled by the refractometer and assures fully automated measurements of the samples placed in the vials of the sample changer for measurements.

Multiparameter measurements

Measuring combination of an automatic refractometer and a density meter as widely used in the flavors and fragrances industry

Today’s laboratories do not only want to measure the refractive index of samples, but several additional parameters like density or viscosity to perform efficient quality control. Due to the microprocessor control and a number of interfaces, automatic refractometers are able to communicate with computers or other measuring devices, e.g. density meters, pH meters or viscosity meters, to store refractive index data and density data (and other parameters) into one database.

Software features

Automatic refractometers do not only measure the refractive index, but offer a lot of additional software features, like

  • Instrument settings and configuration via software menu
  • Automatic data recording into a database
  • User-configurable data output
  • Export of measuring data into Microsoft Exceldata sheets
  • Statistical functions
  • Predefined methods for different kinds of applications
  • Automatic checks and adjustments
  • Check if sufficient amount of sample is on the prism
  • Data recording only if the results are plausible

Pharma documentation and validation

Refractometers are often used in pharmaceutical applications for quality control of raw intermediate and final products. The manufacturers of pharmaceuticals have to follow several international regulations like FDA 21 CFR Part 11, GMP, Gamp 5, USP<1058>, which require a lot of documentation work. The manufacturers of automatic refractometers support these users providing instrument software fulfills the requirements of 21 CFR Part 11, with user levels, electronic signature and audit trail. Furthermore, Pharma Validation and Qualification Packages are available containing

  • Qualification Plan (QP)
  • Design Qualification (DQ)
  • Risk Analysis
  • Installation Qualification (IQ)
  • Operational Qualification (OQ)
  • Check List 21 CFR Part 11 / SOP
  • Performance Qualification (PQ)

SCALES USED ARE:

Reference

https://en.wikipedia.org/wiki/Refractometer

www.anton-paar.com

 www.rudolphresearch.com

https://www.coleparmer.com/tech-article/refractometers

Written by Adi Oluwakemi F

[email protected], 08060874724

www.aasnig.com